2022
DOI: 10.12998/wjcc.v10.i33.12352
|View full text |Cite
|
Sign up to set email alerts
|

Pacemaker electrode rupture causes recurrent syncope: A case report

Abstract: BACKGROUND Currently, the implantation of permanent cardiac pacemakers entails mostly subclavian vein puncture, which is relatively simpler and easier to master. However, due to individual differences, some patients carry a narrow space between the clavicle and the first rib. If the range of activity of the upper limb is increased, the friction between the electrode wire and the bone gap leads to the breakage of the electrode wire, which is manifested by poor pacemaker perception and pacing. … Show more