Recovery of spores from environmental surfaces varies due to sampling and analysis methods, spore size and characteristics, surface materials, and environmental conditions. Tests were performed to evaluate a new, validated sponge wipe method using Bacillus atrophaeus spores. Testing evaluated the effects of spore concentration and surface material on recovery efficiency (RE), false-negative rate (FNR), limit of detection (LOD), and their uncertainties. Ceramic tile and stainless steel had the highest mean RE values (48.9 and 48.1%, respectively). Faux leather, vinyl tile, and painted wood had mean RE values of 30.3, 25.6, and 25.5, respectively, while plastic had the lowest mean RE (9.8%). Results show roughly linear dependences of RE and FNR on surface roughness, with smoother surfaces resulting in higher mean REs and lower FNRs. REs were not influenced by the low spore concentrations tested ( -an integrated network of state and local public health, federal, military, and international laboratories that can respond to bioterrorism, chemical terrorism, and other public health emergencies. However, results from surface samples were inconsistent. In some cases, contamination at a given location within a facility was not detected with initial samples, and subsequent samples were required to detect the contamination (19). A Government Accountability Office (GAO) investigation following the 2001 anthrax incident concluded that validated sampling methods and statistical sampling designs were needed to provide confidence that there is no contamination when all sample results are negative (17, 18). This conclusion strongly reinforces the need for validated sampling methods to effectively respond to biothreats and ensure public safety.Following the 2001 anthrax incident, several research teams developed and investigated (in laboratory studies) the performance of sampling methods using swab, wipe, and vacuum collection devices for Bacillus anthracis or surrogate contaminants on different surfaces (1,2,3,4,5,6,7,13,14,15,16,22,23,25,28,34,38,40,44,45). In addition, the CDC has conducted formal validation studies on two methods for sampling nonporous surfaces: macrofoam swabs (23) and cellulose sponge wipes (39).A review of the cited laboratory studies identified numerous gaps in the data on method performance (30). For example, none of the studies quantified the false-negative rate (FNR) for the sampling and analysis methods investigated. The term "false negative" refers to a failure to detect contamination from a sample collected at a contaminated location. A false negative can occur because of inefficiencies (i.e., biases) and uncertainties (i.e., imprecision) at any step of the sampling and analysis process (sample collection, storage/transportation, processing/extraction, and analytical). False negatives can occur during preliminary screening or characterization sampling at low contamination levels and during clearance sampling following the decontamination process. A better understanding of FNRs and how they are infl...