2017
DOI: 10.1007/s00540-017-2361-z
|View full text |Cite
|
Sign up to set email alerts
|

Paravertebral block catheter breakage by electrocautery during thoracic surgery

Abstract: Advantages of thoracic paravertebral analgesia (TPA) include placement of the catheter closer to the surgical field; however, the catheter can become damaged during the operation. We experienced a case of intraoperative TPA catheter breakage that prompted us to perform an experiment to investigate possible causes. A 50-year-old male underwent a thoracoscopic lower lobectomy under general anesthesia with TPA via an intercostal approach. Following surgery, it was discovered that the catheter had become occluded,… Show more