2018
DOI: 10.14316/pmp.2018.29.2.59
|View full text |Cite
|
Sign up to set email alerts
|

Patient-Specific Quality Assurance in a Multileaf Collimator-Based CyberKnife System Using the Planar Ion Chamber Array

Abstract: This paper describes the clinical use of the dose verification of multileaf collimator (MLC)-based CyberKnife plans by combining the Octavius 1000SRS detector and water-equivalent RW3 slab phantom. The slab phantom consists of 14 plates, each with a thickness of 10 mm. One plate was modified to support tracking by inserting 14 custom-made fiducials on surface holes positioned at the outer region of 10×10 cm 2 . The fiducial-inserted plate was placed on the 1000SRS detector and three plates were additionally st… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
6
0

Year Published

2019
2019
2025
2025

Publication Types

Select...
5

Relationship

1
4

Authors

Journals

citations
Cited by 5 publications
(6 citation statements)
references
References 13 publications
0
6
0
Order By: Relevance
“…In the low resolution area, filling the remainder of the 11 × 11 cm 2 area, the distance between the chambers is 5 mm. The use of the 1000SRS array for the CyberKnife has been investigated and found sufficiently accurate for patient‐specific QA measurements 4,6 . Our 1000SRS array has a fixed geometry with a slab phantom in which a set of three fiducial markers are embedded.…”
Section: Methodsmentioning
confidence: 99%
See 4 more Smart Citations
“…In the low resolution area, filling the remainder of the 11 × 11 cm 2 area, the distance between the chambers is 5 mm. The use of the 1000SRS array for the CyberKnife has been investigated and found sufficiently accurate for patient‐specific QA measurements 4,6 . Our 1000SRS array has a fixed geometry with a slab phantom in which a set of three fiducial markers are embedded.…”
Section: Methodsmentioning
confidence: 99%
“…The use of the 1000SRS array for the Cyber-Knife has been investigated and found sufficiently accurate for patient-specific QA measurements. 4,6 Our 1000SRS array has a fixed geometry with a slab phantom in which a set of three fiducial markers are embedded. Pre-treatment alignment on the CyberKnife is based on a fiducial match between two stereoscopic x-ray images with digitally reconstructed radiographs based on the treatment planning CT, ensuring optimal alignment between planned and delivered geometry.…”
Section: C | Comparison To Pre-treatment Measurementsmentioning
confidence: 99%
See 3 more Smart Citations