2015
DOI: 10.1002/ccd.25940
|View full text |Cite
|
Sign up to set email alerts
|

Percutaneous closure of patent ductus arteriosus in children using amplatzer duct occluder II: Relationship between PDA type and risk of device protrusion into the descending aorta

Abstract: Device closure of PDA using the ADO II is a safe procedure for chosen types of PDA. We demonstrated a novel technique for objective assessment of device protrusion into the descending aorta based on measurable parameters. ADOII device closure of non-conical PDAs warrants closer follow ups.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1

Citation Types

0
23
0

Year Published

2017
2017
2022
2022

Publication Types

Select...
7

Relationship

0
7

Authors

Journals

citations
Cited by 14 publications
(23 citation statements)
references
References 10 publications
0
23
0
Order By: Relevance
“…[212223] The Liddy group reported an occlusion rate of 98% after 6 months. [22] In our study, we noted a lower immediate occlusion rate (85.7%), and this rate remained the same after 3 months.…”
Section: Discussionmentioning
confidence: 99%
See 2 more Smart Citations
“…[212223] The Liddy group reported an occlusion rate of 98% after 6 months. [22] In our study, we noted a lower immediate occlusion rate (85.7%), and this rate remained the same after 3 months.…”
Section: Discussionmentioning
confidence: 99%
“…group did not report hemolysis among their series. [23] A possible explanation for the hemolysis noted with ADO II devices is the presence of residual PDA flow, exaggerated by the use of heparin and streptokinase for the treatment of arterial thrombosis.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…In Vietnam, an experienced pediatric cardiologist empirically noticed that tubulartype PDA was more frequently seen in PDA associated with CRS than in general PDA without CRS (Do TN, personal communication). Transcatheter closure of tubular-type PDA has difficulty in stabilizing the prosthesis due to lack of a sufficient ampulla; there is the risk of displacement, embolization, or aortic protrusion [62,63]. To clarify the cardiologist's notion and investigate morphological and hemodynamic characteristics of PDA associated with CRS, a retrospective survey of 108 children with CRS and 290 children with PDA but without CRS was conducted in Ho Chi Minh City (Toizumi et al, under review).…”
Section: Cardiac Diseases Of Congenital Rubella Syndromementioning
confidence: 99%
“…1) In recent years, transcatheter closure of PDA has become the gold standard treatment approach in patients weighing >5 kg. [1][2][3] The Amplatzer Duct Occluder (ADO; St. Jude Medical Inc., Minnesota, USA) device has been demonstrated to be useful for closing usual conical-shaped PDAs (Krichenko A). 4) However, left pulmonary artery stenosis and/or aortic obstruction has been shown to be a major problem in low-weight infants.…”
mentioning
confidence: 99%