Abstract: A laboratory prototype is described of a new ICP-MS system with high mass resolution, based on a double focusing magnetic field analyser, which has been especially designed for application in elemental analysis. The performance characteristics of this instrument are summarized comparing operation in a low resolution mode (R=300) and a high resolution mode (R=3000) with reference to common quadrupole instruments with (nearly) unit resolution. Most of the typical spectral interferences can be resolved. Detection… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

0
64
1

Year Published

2001
2001
2016
2016

Publication Types

Select...
6
1

Relationship

1
6

Authors

Journals

citations
Cited by 98 publications
(65 citation statements)
references
References 17 publications
0
64
1
Order By: Relevance
“…A Finnigan MAT ELEMENT ICP-MS was used (12,13,17). ICP conditions: outer gas flow rate 14 L min Ϫ1 , auxiliary gas flow rate 0.8 L min Ϫ1 , forward power 1.25 kW, sampling position 10 mm from load coil, on center, unshielded load coil.…”
Section: Methodsmentioning
confidence: 99%
See 2 more Smart Citations
“…A Finnigan MAT ELEMENT ICP-MS was used (12,13,17). ICP conditions: outer gas flow rate 14 L min Ϫ1 , auxiliary gas flow rate 0.8 L min Ϫ1 , forward power 1.25 kW, sampling position 10 mm from load coil, on center, unshielded load coil.…”
Section: Methodsmentioning
confidence: 99%
“…For some important elements, such as Fe, S, and P, ICP-MS with a quadrupole mass analyzer suffers from polyatomic interference problems. The use of a magnetic sector mass analyzer can overcome many of these spectral interference problems (11)(12)(13). Compared to quadrupoles, sector instruments also provide extremely low instrumental background, improved sensitivity in low resolution for isotopes that are not prone to spectral interference, and very good detection limits (0.1 to 1 ppt) for nearly all elements.…”
mentioning
confidence: 98%
See 1 more Smart Citation
“…Low mass fractions of "metallic" impurities were preferably determined by ICP mass spectrometry using a high resolution sector field mass spectrometer (HR ICP MS). The advantages of this method for the described task are: its broad element coverage with fast multi-element capability, its high selectivity in high resolution mode [6][7][8] and its high sensitivity when low resolution mode was used.…”
Section: Methodsmentioning
confidence: 99%
“…The resulting instrument, the Element, was presented in 1993 and was introduced to the market in 1994. [62][63][64][65] An improved version of this instrument, the Element 2, was launched in 1998.…”
Section: Single Collector Sector Field Instruments (Icp-sfms)mentioning
confidence: 99%