Background: Outcomes in patients with relatively high His-bundle (HB) capture thresholds at implantation are unknown. This study aimed to compare changes in the HB capture threshold and prognosis between patients with a relatively high threshold and those with a low threshold.
Methods and Results:Forty-nine patients who underwent permanent HB pacing (HBP) were divided into two groups: low (<1.25 V at 1.0 ms; n=35) and high (1.25-2.49 V; n=14) baseline HB capture threshold groups. The HB capture threshold was evaluated at implantation, and after 1 week, 1, 3, and 6 months, and every 6 months thereafter. HB capture threshold rise was defined as threshold rise ≥1.0 V at 1.0 ms compared with implantation measures. We compared outcomes between the groups. During a mean followup period of 34.6 months, the high-threshold group showed a trend toward a higher incidence of HB capture threshold of ≥2.5 V (50% vs. 14%; P=0.023), HBP abandonment (29% vs. 8.6%; P=0.091), lead revision (21% vs. 2.9%; P=0.065), and clinical events (allcause death, heart failure hospitalization, and new-onset or progression of atrial fibrillation; 50% vs. 23%; P=0.089) than the lowthreshold group. A baseline HB capture threshold of ≥1.25V was an independent predictor of clinical events.
Conclusions:A relatively high HB capture threshold is associated with increased risk of HBP abandonment, lead revision, and poor clinical outcomes.