We have developed a compact photoelectron imaging facility, including an anion source with dissociative photoelectron attachment to molecules, a linear time-of-flight mass spectrometry (TOFMS), and an orthogonal high-resolution threshold photoelectron velocity map imaging spectrometer (VMI). Intense and cold cluster anions were prepared in photoelectronattachment processes upon pulsed UV laser ablation of metal target. Combining this anion source with TOFMS-VMI, the achieved mass resolution is about 200, and the electron kinetic energy resolution is better than 3%, i.e., 30 meV for 1 eV electrons. More importantly, low-energy photoelectron imaging spectra for CH 3 S − and S 2 − at 611.46 nm are obtained. In both cases, the refined electron affinities are determined to be 1.8626±0.0020 eV for CH 3 S and 1.6744±0.0035 eV for S 2 , respectively. Preliminary results suggest that the apparatus is a powerful tool for estimating precise electron affinities values from threshold photoelectron imaging spectroscopy.