2021
DOI: 10.1136/bmjinnov-2020-000536
|View full text |Cite
|
Sign up to set email alerts
|

Pneumatic femoral broaching in direct anterior approach hip arthroplasty

Abstract: AimThe aim of this study is to report the safety and efficacy of the Woodpecker pneumatic broaching system in direct anterior hip arthroplasty.Methods649 primary elective anterior bikini total hip arthroplasties (THA) using Woodpecker broaching over a 5-year period were included. Patients undergoing a THA through a different surgical approach, revision THA or arthroplasties for hip fractures were excluded (n=219). Preoperative and postoperative Harris Hip Scores (HHS) and postoperative radiographs were analyse… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
1
0

Year Published

2021
2021
2021
2021

Publication Types

Select...
2

Relationship

1
1

Authors

Journals

citations
Cited by 2 publications
(1 citation statement)
references
References 19 publications
0
1
0
Order By: Relevance
“…We did not come across any case of femoral nerve injury in our series. In addition, the pneumatic broaching system minimises the time the operative leg is placed in extension, adduction and external rotation thereby minimising stretching of the femoral nerve [27].…”
Section: Discussionmentioning
confidence: 99%
“…We did not come across any case of femoral nerve injury in our series. In addition, the pneumatic broaching system minimises the time the operative leg is placed in extension, adduction and external rotation thereby minimising stretching of the femoral nerve [27].…”
Section: Discussionmentioning
confidence: 99%