Abstract:A power meter was constructed to measure the power being deposited in transient laser-induced plasmas (LIP). The design rejects the radiation scattered from the laser beam by the LIP and that transmitted past the LIP. The power meter is constructed from a small metal-capped flask placed in an evacuated chamber. The steady-state heat production in the flask is determined from measurements of the temperature gradient on a thermal resistance connecting the top of the flask with a heat sink. The design worked well… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.