Problems with capnography at high altitude*

Abstract: SummaryThe aim of this study was to investigate why capnographs malfunction at high altitude, in order to suggest ways of improving instrument reliability. Four machines were decompressed to simulated high altitude and measurements taken. The study showed that the reduced air density at altitude has effects upon the pump, causing difficulty in entraining gas, and that high altitude has additional effects upon calibration of the instruments. There is ongoing interest in the effects of carbon dioxide (CO 2 ) on … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

0
8
0

Year Published

2005
2005
2021
2021

Publication Types

Select...
4
3

Relationship

0
7

Authors

Journals

citations
Cited by 17 publications
(8 citation statements)
references
References 6 publications
0
8
0
Order By: Relevance
“…The Cosmed K4b 2 was chosen for its portability and performance at high altitude (6), which has been confirmed subsequently by the authors (36). Gases were collected via a tight-fitting Cosmed-modified Hans Rudolph facemask.…”
Section: Subjectsmentioning
confidence: 99%
“…The Cosmed K4b 2 was chosen for its portability and performance at high altitude (6), which has been confirmed subsequently by the authors (36). Gases were collected via a tight-fitting Cosmed-modified Hans Rudolph facemask.…”
Section: Subjectsmentioning
confidence: 99%
“…correcting for the transport delay between the turbine and gas analyzers and for the analyzer rise time (Beaver et al, 1973). The accuracy of the telemetric system has been previously established (Palange et al, 2000;Pattinson et al, 2004). Heart rate was calculated beat-by-beat, and SpO 2 was monitored non-invasively using finger pulse oximetry (Masimo Rad-5, Masimo Corporation, Irvine, CA, USA).…”
Section: Equipmentmentioning
confidence: 99%
“…Prior to each test, the zero-load setting on the ergometer was checked and a calibration was performed with a 4 kg weight. Ventilatory and pulmonary gas exchange variables were measured breath-by-breath in all tests using a portable system (K4b 2 , COSMED, Italy), which has previously been used at HA (Pattinson et al, 2004). The system comprised a face mask, analyzer unit (containing O 2 and CO 2 gas analyzers), heart-rate monitor and battery.…”
Section: Equipmentmentioning
confidence: 99%
“…Capnograph malfunction is common at HA, possibly due to reduced gas fl ow rates through the sample chamber, the effect of low barometric pressure on the calibration, or the effect of the reduced barometric pressure on the computer software [86]. Furthermore, signifi cant hyperventilation and low P CO 2 levels may make colorimetric P CO 2 detectors unreliable [87].…”
Section: Anesthetics and Anesthesia Equipment At Hamentioning
confidence: 98%
“…Barometric pressure also affects the function of capnographs, which need to be recalibrated at the ambient pressure [86]. Capnograph malfunction is common at HA, possibly due to reduced gas fl ow rates through the sample chamber, the effect of low barometric pressure on the calibration, or the effect of the reduced barometric pressure on the computer software [86].…”
Section: Anesthetics and Anesthesia Equipment At Hamentioning
confidence: 99%