Profiling of adrenocorticotropic hormone and arginine vasopressin in human pituitary gland and tumor thin tissue sections using droplet-based liquid-microjunction surface-sampling-HPLC–ESI-MS–MS

Abstract: Described here are the results from the profiling of the proteins arginine vasopressin (AVP) and adrenocorticotropic hormone (ACTH) from normal human pituitary gland and pituitary adenoma tissue sections using a fully automated droplet-based liquid microjunction surface sampling-HPLC-ESI-MS/MS system for spatially resolved sampling, HPLC separation, and mass spectral detection. Excellent correlation was found between the protein distribution data obtained with this droplet-based liquid microjunction surface sa… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1

Citation Types

1
24
0

Year Published

2015
2015
2021
2021

Publication Types

Select...
6

Relationship

2
4

Authors

Journals

citations
Cited by 25 publications
(25 citation statements)
references
References 39 publications
(74 reference statements)
1
24
0
Order By: Relevance
“…Typically, pictures of the junction before and after solvent dispense, and of the surface after solvent aspirations, were collected for all extraction cycles. Glass slides were mounted in a 3D‐printed spring‐loaded microtiter‐plate‐size sample holder . The sample holder was then placed into an open tray of the HTS PAL autosampler and the analysis locations in each sample were selected with a laser distance sensor using software developed in house (dropletProbe Premium v1.20).…”
Section: Methodsmentioning
confidence: 99%
See 2 more Smart Citations
“…Typically, pictures of the junction before and after solvent dispense, and of the surface after solvent aspirations, were collected for all extraction cycles. Glass slides were mounted in a 3D‐printed spring‐loaded microtiter‐plate‐size sample holder . The sample holder was then placed into an open tray of the HTS PAL autosampler and the analysis locations in each sample were selected with a laser distance sensor using software developed in house (dropletProbe Premium v1.20).…”
Section: Methodsmentioning
confidence: 99%
“…Liquid extraction‐based surface sampling probes of different types are receiving increased attention due to their enhanced sensitivity compared with other ambient surface sampling techniques . and to their ability to integrate with HPLC/MS systems . These techniques for spatially resolved sampling rely on liquid‐solid extraction of material from the tissue.…”
mentioning
confidence: 99%
See 1 more Smart Citation
“…54 Plug formation mode enables analyte preconcentration and reactive analysis. 51 LMJ-SSP has been used for the analysis of molecules separated on TLC plates, 47 tissue profiling, 55,56 and spatial mapping of metabolites in microbial 57 and fungal communities. 58 …”
Section: Ambient Liquid Extraction Techniquesmentioning
confidence: 99%
“…Since that time multiple continuous flow as well as discrete droplet variations of this liquid microjunction surface sampling and analysis approach have been implemented for numerous applications. Some of these liquid microjunction probes/methods allow ready transfer of the sampled material for post‐sampling processing, such as a high‐performance liquid chromatography (HPLC) separation, prior to mass spectral analysis . Such a configuration enables effective analysis of targeted analytes within complex matrices as well as the ability to differentiate isobaric compounds, thus providing a greater degree of chemical information from a single spot sample than is possible with a direct analysis.…”
mentioning
confidence: 99%