Purpose
To compare the rotational stability of a monofocal and a diffractive multifocal toric intraocular lenses with identical design and material.
Methods
This prospective study enrolled patients who underwent plate-haptic toric IOL (AT TORBI 709 M and AT LISA 909 M) implantation. Propensity score matching (PSM) was performed to balance baseline factors. Follow-up examinations were conducted at 1 hour, 1 day, 3 days, 1 week, 2 weeks, 1 month, and 3 months postoperatively. A linear mixed model of repeated measures was used to investigate the changes in IOL rotation over time. A 2-week timeframe was utilized to assess differences in IOL rotation between the two groups.
Result
After PSM, a total of 126 eyes were selected from each group for further analysis. Postoperatively, the time course of IOL rotation change in the two groups remained consistent, with the greatest rotation occurring between 1 hour and 1 day postoperatively. At the 2-week postoperative mark, the monofocal toric IOL exhibited a higher degree of rotation compared to the multifocal toric IOL (5.40 ± 7.77° vs. 3.53 ± 3.54°, P = 0.015). In LT ≥ 4.5mm and WTW ≥ 11.6mm subgroups, the monofocal toric IOL rotated greater than the multifocal toric IOL (P = 0.026 and P = 0.011, respectively).
Conclusion
The diffractive multifocal toric IOL exhibits superior rotational stability compared to the monofocal toric IOL, especially in subgroups LT ≥ 4.5 mm and WTW ≥ 11.6 mm. Moreover, the time course of IOL rotation change is consistent for both, with the maximum rotation occurring between 1 hour and 1 day postoperatively.