QMS in the Third Stability Zone with a Transverse Magnetic Field Applied

Abstract: We report here a study using a quadrupole mass spectrometer (QMS) in which a static magnetic field is applied transversely to the body of the mass filter operating in stability zone 3. Significant improvement in QMS performance was obtained under certain magnetic field conditions, and these have been explained in terms of our theoretical model. The theoretical approach assumed in the model is that the QMS contains hyperbolic rods as electrodes and that the magnetic field acts over the full length of the mass f… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
4
0

Year Published

2013
2013
2016
2016

Publication Types

Select...
4
1

Relationship

0
5

Authors

Journals

citations
Cited by 5 publications
(4 citation statements)
references
References 20 publications
0
4
0
Order By: Relevance
“…Previously published experimental and theoretical results for QMF operating in stability zone 1 showed: the effect of a transverse magnetic field on the resolution of QMF, the effect of a static magnetic field on the ion trajectories and the effect of an axial magnetic field on the performance of a QMF . For QMF operating in stability zone 3, previous work had shown the effect of transverse magnetic field on resolution . In this paper, therefore, we present new experimental and simulation results explaining the performance of QMF operating in stability zones 1 and 3 under the influence of magnetic field in detail.…”
Section: Introductionmentioning
confidence: 82%
See 2 more Smart Citations
“…Previously published experimental and theoretical results for QMF operating in stability zone 1 showed: the effect of a transverse magnetic field on the resolution of QMF, the effect of a static magnetic field on the ion trajectories and the effect of an axial magnetic field on the performance of a QMF . For QMF operating in stability zone 3, previous work had shown the effect of transverse magnetic field on resolution . In this paper, therefore, we present new experimental and simulation results explaining the performance of QMF operating in stability zones 1 and 3 under the influence of magnetic field in detail.…”
Section: Introductionmentioning
confidence: 82%
“…An option of introducing magnetic in all the three directions (B x , B y , B z ) , along the whole length of the mass filter, is also incorporated. The original model has been upgraded to provide: improved sensitivity (using a larger number of steps along the mass scale), to include large number of ions at each point on the mass scale and to simulate a QMF with a very large number of rf cycles.…”
Section: Softwarementioning
confidence: 99%
See 1 more Smart Citation
“…The application of a static axial magnetic field along the mass filter increases the resolution of a quadrupole mass filter with hyperbolic-shaped electrodes [17]. Quadrupole mass filter with hyperbolic electrodes operated in zone 3 of the Mathieu stability diagram could achieve the resolution of the low mass isotopes such as HT-D2, separated by 0.004125 u [18,19]. The implementation of RGA concerns a very wide application range.…”
Section: B the Chosen Mass Spectrometermentioning
confidence: 99%