, Clin. Vaccine Immunol. 13:541-546, 2006), we investigated the agreement between a validated anthrax protective antigen (PA) immunoglobulin G (IgG) enzyme-linked immunosorbent assay (ELISA) and the LFA for 335 unvaccinated and vaccinated subjects. We also investigated the performance of the LFA under the following conditions: thermal shock (i.e., thermal cycling between temperature extremes), high temperature/high relative humidity, high temperature/low relative humidity, and low temperature/low relative humidity. With the anti-PA ELISA used as a standard, the LFA was shown to be optimally diagnostic at 11 g/ml anti-PA-specific IgG. At this concentration, the LFA specificity and sensitivity were 98% (95% confidence interval [CI], 97% to 100%) and 92% (CI, 88% to 97%), respectively. Receiver operating characteristic curve analysis yielded an area under the curve value of 0.988 (CI, 0.976 to 1.00), suggesting that the LFA is an extremely accurate diagnostic test. For <4 or >50 g/ml PA-specific IgG, the LFA results for each environmental condition were identical to those obtained in the laboratory. These data indicate that this rapid point-of-care test would be a feasible tool in monitoring the serological antibody responses of individuals that have been vaccinated against anthrax.