Adoption of interventional endoscopic procedures is increasing with increasing prevalence of diseases. However, medical radiation exposure is concerning; therefore, radiation protection for medical staff is important. However, there is limited information on the usefulness of an additional lead shielding device during interventional endoscopic procedures. Therefore, we aimed to determine whether an additional lead shielding device protects medical staff from radiation.
An X-ray unit (CUREVISTA; Hitachi Medical Systems, Tokyo, Japan) with an over-couch X-ray system was used. Fluoroscopy-associated scattered radiation was measured using a water phantom placed at the locations of the endoscopist, assistant, nurse, and clinical engineer. For each location, measurements were performed at the gonad and thyroid gland/eye levels. Comparisons were performed between with and without the additional lead shielding device and with and without a gap in the shielding device. Additionally, a clinical study was performed with 27 endoscopic retrograde cholangiopancreatography procedures.
The scattered radiation dose was lower with than without additional lead shielding at all medical staff locations and decreased by 84.7%, 82.8%, 78.2%, and 83.7%, respectively, at the gonad level and by 89.2%, 86.4%, 91.2%, and 87.0%, respectively, at the thyroid gland/eye level. Additionally, the scattered radiation dose was lower without than with a gap in the shielding device at all locations.
An additional lead shielding device could protect medical staff from radiation during interventional endoscopic procedures. However, gaps in protective equipment reduce effectiveness and should be eliminated.