Applications of lasers in phototherapy have been the trend for the last few decades. The photodynamic therapy process normally depends on photosensitizers and laser beams. Through this study, indocyanine green has been used as a photosensitizer, which is normally activated using laser lines between 750 and 805 nm. The activity of the indocyanine green to do fluorescence by other pulsed laser sources has been tested by fluorescence technique, and it has been proven that the laser lines at 810, 940, and 980nm are able to excite the indocyanine green with different extents. The indocyanine green activation has been tested by several laser lines (810, 940, and 980 nm) commonly used as surgical lasers. The generated oxygen has been measured after irradiating the indocyanine green with the different laser lines. A comparison has been made between laser irradiation as a pinpoint and a broad beam. It is found that the wide beam is more effective in activating oxygen production. In the end, it is concluded that lines 810 and 940nm were effective in activating the used dye, while the 980nm activity did not show enough efficiency.