2024
DOI: 10.1111/aor.14903
|View full text |Cite
|
Sign up to set email alerts
|

Real‐time physiological environment emulation for the Istanbul heart ventricular assist device via acausal cardiovascular modeling

Hammad Ur Rahman,
Khunsha Mehmood,
Farouk Abdulhamid
et al.

Abstract: Background and ObjectivesThe cost and complexity associated with animal testing are significantly reduced by using mock circulatory loops prior. Novel mock circulatory loops allow us to test biomedical devices preclinically due to their flexibility, scalability, and cost‐effectiveness. The presented work describes the development of a hardware‐in‐the‐loop platform to emulate human physiology for the Istanbul Heart (iHeart‐II) LVAD.MethodsA closed‐loop system is developed whereby the effect of the LVAD on the h… Show more