Reducing measurement uncertainty in 63Ni measurements in reactor coolant water with high 60Co activities

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
5
0

Year Published

2016
2016
2023
2023

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 5 publications
(5 citation statements)
references
References 8 publications
0
5
0
Order By: Relevance
“…The 63 Ni beta spectrum peaks at channel region 250-300 and 60 Co beta spectrum peaks at channel region 500-550. The measurement time was 360 min these samples, then it might have been possible to achieve a negligible activity level of 60 Co in the 63 Ni fractions [23].…”
Section: Discussionmentioning
confidence: 99%
“…The 63 Ni beta spectrum peaks at channel region 250-300 and 60 Co beta spectrum peaks at channel region 500-550. The measurement time was 360 min these samples, then it might have been possible to achieve a negligible activity level of 60 Co in the 63 Ni fractions [23].…”
Section: Discussionmentioning
confidence: 99%
“…The preliminary plan was to dissolve the steel samples in concentrated acids and, after hydroxide precipitation of investigated metals, separate Fe and Ni from each other and disturbing matrix with anion exchange. Then Ni fractions from anion exchange would be further purified from Co either by co-precipitating Ni with DMG (traditional method) [8,11] or by extraction chromatographic column separation with Ni resin (a newer method) [12][13][14]. Actually, the both methods are based on the complexation of Ni by DMG but the procedures are different in practice, direct mixing of sample and DMG solutions in a beaker vs. column separation.…”
Section: Testing Of Methods 2 In Determination Of 55 Fe and 63 Nimentioning
confidence: 99%
“…In order to verify the effectiveness of the Ni purification from 60 Co, the eluate was measured using gamma spectrometry. The results showed significant presence of 60 Co and thus, the Ni-resin purification was repeated according to Eriksson et al [13].…”
Section: Determination Of 14 C 55 Fe and 63 Ni Using Methods 11mentioning
confidence: 99%
“…However, the change of two subsequent 10 g AG resins to one 30 g AG resin was not successful to increase the Fe yield in the purified fractions as the yields lowered from 48 to 80% to 13-19%. As discussed in previous publications [12,14,22,27], purification of Ni from Co can be challenging. The yield results for Ni in inactive concrete subsamples show that about 90% of the Ni was carried over to the final purified Ni fraction.…”
Section: Testing Of Methods 1 In Inactive Concrete and The Radiochemical Analysis Of 41 Ca 55 Fe And 63 Ni In Activated Concretementioning
confidence: 99%
“…CoreF correction is a mathematical corrective function for deviations of TDCR at higher quench levels. [2,[22][23][24]28]. Three parallel subsamples of 0.5 g inactive concrete were used for experimenting the initial Method 2.…”
Section: Testing Of Methods 1 In Inactive Concrete and The Radiochemical Analysis Of 41 Ca 55 Fe And 63 Ni In Activated Concretementioning
confidence: 99%