2022
DOI: 10.1002/acm2.13861
|View full text |Cite
|
Sign up to set email alerts
|

Reducing PTV margins for prostate SBRT with motion compensation and gating techniques

Abstract: The purpose of this study is to investigate the dosimetric accuracy of prostate SBRT when motion is considered. To account for target movement, motion compensation and gating techniques were investigated with PTV margins reduced to 2 mm. To allow for dosimetric measurements a Delta4 phantom, Gafchromic film, and Hexamotion motion platform were utilized. Four motion files were utilized that represent a range of motions. Analysis of measured prostate motions for fifteen patients was performed to ensure detected … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

0
3
0

Year Published

2022
2022
2024
2024

Publication Types

Select...
5
1

Relationship

0
6

Authors

Journals

citations
Cited by 6 publications
(3 citation statements)
references
References 34 publications
0
3
0
Order By: Relevance
“…Real-time tracking of the prostate via surrogate fiducial markers may permit a similar reduction in PTV margins ( 30 ). However, fiducial marker implantation is an invasive procedure accompanied by a slightly increased risk for infection or even fiducial migration in rare cases which might lead to impaired tracking ( 31 ).…”
Section: Discussionmentioning
confidence: 99%
“…Real-time tracking of the prostate via surrogate fiducial markers may permit a similar reduction in PTV margins ( 30 ). However, fiducial marker implantation is an invasive procedure accompanied by a slightly increased risk for infection or even fiducial migration in rare cases which might lead to impaired tracking ( 31 ).…”
Section: Discussionmentioning
confidence: 99%
“…With the introduction of the Synchrony system, tomotherapy can be used to track SBRT [ 9 , 10 ]. By using the system, the 3D fiducial position is calculated based on successive 2D kV radiographs, and the translational motion can be compensated by jaw sweeping in the longitudinal direction and MLC shifting in the lateral and vertical directions.…”
Section: Discussionmentioning
confidence: 99%
“…Minor deviations may lead to undercoverage of target volumes and higher toxicity rates ( 10 ). Several studies have attempted to assess the optimal target margin in prostate SBRT using image-guided RT (IGRT); however, they obtained limited information on prostate movement owing to the lack of full-time continuous monitoring or the use of X-ray-based image guidance ( 11 , 12 ).…”
Section: Introductionmentioning
confidence: 99%