Abstract:Precise fluid administration is important to prevent hypo- or hypervolemia. However, the accuracy of scales marked on intravenous (IV) fluid plastic bags had remained unknown. Ten 1 L sized IV crystalloids were prepared from each of three manufacturers (H, J, and D). At each scale, the actual volume of the IV fluid was measured. Differences with the measured volumes for each scale were investigated between the three manufacturers. All initial total volume was greater than 1 L. Except for the full-filled level,… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.