Reliability of the point-of care analyzer “StatStrip® Xpress™” for measurement of fetal blood lactate

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

1
4
0

Year Published

2021
2021
2022
2022

Publication Types

Select...
3

Relationship

1
2

Authors

Journals

citations
Cited by 3 publications
(5 citation statements)
references
References 27 publications
1
4
0
Order By: Relevance
“…The POC CUAV lactate concentration was strongly and significantly correlated with CUAV lactate concentration measured on the reference analyzer ( R 2 = 0.928, F = 1.26 × 10 −56 , Figure 3). Like Iorizzo et al, 17 our POC‐derived values were generally lower than those produced by the reference analyzer. The mean difference between CUAV lactate analyzed at the POC versus the reference analyzer was 0.299 mmol/L (95% CI, 0.220–0.376), which is considered clinically acceptable.…”
Section: Resultssupporting
confidence: 67%
See 2 more Smart Citations
“…The POC CUAV lactate concentration was strongly and significantly correlated with CUAV lactate concentration measured on the reference analyzer ( R 2 = 0.928, F = 1.26 × 10 −56 , Figure 3). Like Iorizzo et al, 17 our POC‐derived values were generally lower than those produced by the reference analyzer. The mean difference between CUAV lactate analyzed at the POC versus the reference analyzer was 0.299 mmol/L (95% CI, 0.220–0.376), which is considered clinically acceptable.…”
Section: Resultssupporting
confidence: 67%
“…The StatStripXpress (SSX; Nova Biomedical, Waltham, MA, USA) was used to assess CUAV lactate concentrations at the POC. Lactate concentrations in UA and UV samples analyzed on the SSX have been shown to correlate closely with those analyzed on reference methods 17 . The present study seeks to determine whether this relationship persists when CUAV samples are used.…”
Section: Methodsmentioning
confidence: 89%
See 1 more Smart Citation
“…6,10,11 SSLX has been shown to perform with good accuracy and agreement to the stationary blood gas machines. [11][12][13][14][15] It is important to recognise that lactate values vary depending on the POC-lactate-meter used. 1,6 As LP is no longer produced, there is an urgent need for evaluation of a new POC-lactatemeter, with assessment of a safe interventional cutoff to rule out fetal acidosis.…”
Section: Introductionmentioning
confidence: 99%
“…The performance and accuracy in fetal and adult blood for several POC‐lactate‐meters have been published 6,10,11 . SSLX has been shown to perform with good accuracy and agreement to the stationary blood gas machines 11–15 . It is important to recognise that lactate values vary depending on the POC‐lactate‐meter used 1,6 .…”
Section: Introductionmentioning
confidence: 99%