Renal Cyst Pseudoenhancement at Multidetector CT: What Are the Effects of Number of Detectors and Peak Tube Voltage?

Abstract: Purpose:To determine the effect of the number of detectors and peak tube voltage on renal cyst pseudoenhancement in a phantom model. Materials and Methods:This study on computed tomographic (CT) phantoms did not require institutional review board approval. The renal compartments of a CT phantom were filled with iodinated contrast material diluted to attain attenuations of 40, 140, and 240 HU. Saline-filled cylinders simulating cysts of varying diameters (range, 0.7-3.0 cm) were serially suspended in the renal … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
25
0

Year Published

2011
2011
2021
2021

Publication Types

Select...
4
4

Relationship

0
8

Authors

Journals

citations
Cited by 64 publications
(25 citation statements)
references
References 17 publications
0
25
0
Order By: Relevance
“…13,18 Although recent CT technologies have improved, a number of artefacts can lead to pseudoenhancement. 19 In our study with a cut-off value of 20 HU for contrast enhancement, two and nine simple renal cysts demonstrated pseudoenhancement on linearly blended images from the CMP and LNP, respectively.…”
Section: Discussionmentioning
confidence: 61%
“…13,18 Although recent CT technologies have improved, a number of artefacts can lead to pseudoenhancement. 19 In our study with a cut-off value of 20 HU for contrast enhancement, two and nine simple renal cysts demonstrated pseudoenhancement on linearly blended images from the CMP and LNP, respectively.…”
Section: Discussionmentioning
confidence: 61%
“…To produce a difference of 8 HU (0.5 mg Au ml -1 ) with 0.03% id g -1 delivery, 1.7 g of injected gold would be required. However, newer dual-energy and multihead CTs are able to reduce the detection levels further [45,46]. X-ray imaging equipment that would exploit gold's favourable absorption edges would not only improve sensitivity but also reduce patient dose.…”
Section: Micro-ct Of Her2-targeted Gold Nanoparticlesmentioning
confidence: 99%
“…Clinically, increased attenuation of small simple renal cysts on contrast-enhanced CT is reported as 'pseudoenhancement;' however, enhancement of a renal mass on contrast-enhanced CT is an essential criterion for distinguishing between tumors and benign cysts [5][6][7][8][9][10][11][12][13]. The present results show the effect of the shape of surrounding tissue on the CT number of the lesion of interest.…”
Section: Discussionmentioning
confidence: 62%
“…The very high linearity observed for all imaging methods and phantoms supports the assumption that focal artifacts do not cause slope inconsistency in conventional imaging; rather, a more diffuse variation of CT numbers exists that can be eliminated by using dualenergy CT imaging. In previously published papers, beam hardening has been frequently mentioned as a cause of pseudoenhancement [5][6][7][8][9][10][11][12][13]. When lower energy photons of the beam are preferentially attenuated by higher density material than water, and images are reconstructed with an algorithm that compensates the excessive attenuation, beam hardening occurs.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation