Reproducibility of optic disc measurements acquired with spectral‐domain optical coherence tomography

Abstract: Purpose To compare the reproducibility of peripapillary retinal nerve fiber layer (RNFL) measurements of 2 optical coherence tomography (OCT) systems. Methods Fifty‐seven healthy individuals were prospectively selected. Only one eye was randomly chosen. All participants had intraocular pressure less than 21 mmHg and normal standard automated perimetry. Peripapillary RNFL thicknesses were measured with the Cirrus (Carl Zeiss Meditec, Dublin, Ca) and Spectralis (Heidelberg Engineering, Heidelberg, Germany) OCTs… Show more