2020
DOI: 10.1111/aor.13790
|View full text |Cite
|
Sign up to set email alerts
|

Research of flow dynamics and occlusion condition in roller pump systems used for ventricular assist

Abstract: Roller pumps have been widely used in the ventricular assist field for many years, while the significant hemolysis caused by its mechanical stress is still a fundamental problem. Although the usual under‐occlusion setting was considered as an effective method to reduce the hemolysis rate, its nonocclusive condition of the whole process may cause serious backflow results, which exactly places many restrictions on this method. In this study, the simulation experiments based on computational fluid dynamics (CFD) … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1

Citation Types

0
6
0

Year Published

2021
2021
2024
2024

Publication Types

Select...
5
1

Relationship

1
5

Authors

Journals

citations
Cited by 7 publications
(6 citation statements)
references
References 27 publications
0
6
0
Order By: Relevance
“…Similar trends were predicted elsewhere. [7][8][9][10] More specifically, the outlet flow rate starts being reduced from φ = 130°(Tube B) or 150°(Tube A and C) till φ = 50°(Tube A) or 70°(Tube B and C). Higher deceleration rates occur in the interval φ = 0°to φ = 50°for Tube A and B, and from φ = 30°to φ = 70°for Tube C. The higher flow rate reduction is attributed to the increase of the tube volume during the disengagement phase, so that eventually a smaller fluid mass is discharged at the pump outlet.…”
Section: Flow Ratementioning
confidence: 99%
See 1 more Smart Citation
“…Similar trends were predicted elsewhere. [7][8][9][10] More specifically, the outlet flow rate starts being reduced from φ = 130°(Tube B) or 150°(Tube A and C) till φ = 50°(Tube A) or 70°(Tube B and C). Higher deceleration rates occur in the interval φ = 0°to φ = 50°for Tube A and B, and from φ = 30°to φ = 70°for Tube C. The higher flow rate reduction is attributed to the increase of the tube volume during the disengagement phase, so that eventually a smaller fluid mass is discharged at the pump outlet.…”
Section: Flow Ratementioning
confidence: 99%
“…The issue of blood damage has been studied by several groups using computational fluid mechanics (CFD) and making the assumption that the flow is two dimensional. [7][8][9] The CFD treatment of the flow in a peristaltic pump needs a lot of computational resources in order to capture the details of the three dimensional (3D) flow field, and in particular, the region around the moving stenosis. Moreover, it has to be stressed that there is no CFD code nowadays that can solve the flow if the tube is deeply occluded due to singularities which appear when the computational cells touch each other at the occluded part of the tube.…”
Section: Introductionmentioning
confidence: 99%
“…Related models of hemodynamics have achieved great effects in a variety of applications, including cardiac output (CO) monitoring, 5,6 aneurysm diagnosis, 7 calculation of coronary fractional flow reserve (FFR), 8 and evaluation of artificial assistive devices. 9 It is believed that using these models can help improve the prediction of cardiovascular activities, and more importantly, it provides a convincing mechanism explanation for this prediction. 10 Especially, hemodynamic mathematical modeling usually contains three different scales, which have led to three typical models respectively, including 0-dimensional (0-D) model, 11 1-dimensional (1-D) model, 12 and 3-dimensional (3-D) model.…”
Section: Introductionmentioning
confidence: 99%
“…Among these models, cardiovascular hemodynamic modeling has become one of the most active areas, considering its close relationship with clinical routine monitoring. Related models of hemodynamics have achieved great effects in a variety of applications, including cardiac output (CO) monitoring, 5,6 aneurysm diagnosis, 7 calculation of coronary fractional flow reserve (FFR), 8 and evaluation of artificial assistive devices 9 . It is believed that using these models can help improve the prediction of cardiovascular activities, and more importantly, it provides a convincing mechanism explanation for this prediction 10 .…”
Section: Introductionmentioning
confidence: 99%
“…A significant variable, possibly impacting this phenomenon, is the degree of pump occlusion. Despite its known association with haemolysis, inadequate occlusion (both tight occlusive and non-occlusive settings) remains an unaddressed issue, with no international standards for adjusting occlusiveness [ 3 ]. The actual flow delivered by the master RP is a function of several aspects of CPB management, including device-related, disposable-related, patient-related, or procedure-related factors, such as pump model, tubing material (silicone or PVC) and diameter, temperature, blood viscosity, oxygenator model, design and pressure drop, the use of vacuum-assisted venous drainage and the choice between continuous or pulsatile flow.…”
mentioning