The aim: To explore the effect of different bar clip materials (plastic, poly ether ether ketone (PEEK), and zirconia) on retention force with PEEK bar in implant supported mandibular overdenture using the universal testing machine.
Materials and methods:A heat-cured acrylic educational model was constructed and to implants were inserted at the canine region bilaterally. A milled PEEK bar was constructed and screwed to the multiunit abutment which were screwed to the implants. Three different clip materials groups (plastic, PEEK, and zirconia) were constructed. The clips were picked up in the intaglio surface of each denture. Three wrought wires were attached to the overdenture polished surface and connected to each other at the overdenture geometric center. The retention force of each attachment was measured at T0 (insertion) followed by T1 at 360 (3 months), T2 at 720 (6 months), T3 at 1440 (one year), T4 at 2880 (2 years) and T5 at 4320 (3 years) insertion and removal cycles simulating 36 months of usage.Results: A statistically significant difference was found between clip materials at all insertion and removal cycles. The final mean retention force of the zirconia clip group was significantly lower compared to the plastic and PEEK clip groups at T5 simulating 3-year usage.
Conclusion:Within the limitations of this study, it can be concluded that the PEEK and plastic clip materials showed significantly higher retention values compared to zirconia clip materials when used with PEEK bar-retained mandibular implant-supported overdentures after a threeyear simulation of overdenture use. Although the retention loss was unavoidable in the three clip materials, the rate of retention loss differed depending on the clip material.