RF-SQUID to DC-SQUID upgrade of a 28-channel magnetoencephalography (MEG) system

Abstract: A Dornier 28-channel magnetoencephalography (MEG) system equipped with RF-SQUIDs and control electronics made by CTF (Port Coquitlam, BC, Canada) was upgraded by replacing the RF-SQUIDs with DC-SQUIDs developed at the University of Jena, Germany. New preamplifiers were designed that emulate the function of the system's original RF amplifiers. This allowed the continuing use of the system's control electronics without any modification. Retaining these control devices was instrumental for preserving the system's… Show more