In this study, polyamine (pentaethylene hexamine) functionalized silica gel (PA-SG) was prepared as a novel sorbent material for on-line preconcentration of Cd(II), Cu(II), and Ni(II) from complex matrices. The PA-SG was characterized by means of elemental analysis and FT-IR spectroscopic techniques. The effective parameters for on-line preconcentration and separation of Cd(II), Cu(II), Ni(II), and FAAS detection were optimized by investigating the effects of sample pH, type, and concentration of eluent, flow rate of sample and eluent, loop volume, and matrix ions. The linear range of the proposed method was between10-200 µg L −1 for each element, with enrichment factors of 39.9, 30.4, and 43.6 and detection limits of 0.19, 0.73, and 0.91 µg L −1 for Cd(II),