Electrospray ionization
mass spectrometry (ESI-MS) is widely used
in drug development, therapeutic drug monitoring, and other fields.
However, unstable mass spectral signals, especially during the initial
stages of instrument operation, plague analysts. Generally, in quantitative
experiments, the stability of response can be achieved by running
the analytical system for some time. However, the equilibration time
required for the responses of different compounds to stabilize has
been elusive. To investigate the response stability of the ESI-MS
system, 72 compounds with different physicochemical properties were
employed on three systems, and flow injection analysis was performed
in positive ion mode. With the use of 5.00% (response stable factor,
RSF) as the stability limit, about 80% of the compounds were stable
within 60 min. Under a 2.00% criterion, the stabilization time was
significantly longer. The stabilization time varies with different
instruments and physicochemical properties of the compounds. When
positive ion detection is performed in an acidic mobile phase, the
octanol–water partition coefficient (Log P), molecular weight, and molar volume can all affect the time required
to stabilize the response. In general, it is necessary to balance
the ESI-MS system for an appropriate time before sample detection,
especially for the analysis of compounds with strong hydrophilicity,
small molecular weight, or small molar volume under the conditions
above.