Simulation of the Expected Performance of a Seamless Scanner for Brain PET Based on Highly Pixelated CdTe Detectors

Abstract: The aim of this work is the evaluation of the design for a nonconventional PET scanner, the voxel imaging PET (VIP), based on pixelated room-temperature CdTe detectors yielding a true 3-D impact point with a density of 450 channels/cm(3), for a total 6 336 000 channels in a seamless ring shaped volume. The system is simulated and evaluated following the prescriptions of the NEMA NU 2-2001 and the NEMA NU 4-2008 standards. Results show that the excellent energy resolution of the CdTe detectors (1.6% for 511 keV… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

1
26
0

Year Published

2014
2014
2022
2022

Publication Types

Select...
9
1

Relationship

2
8

Authors

Journals

citations
Cited by 19 publications
(27 citation statements)
references
References 13 publications
1
26
0
Order By: Relevance
“…As expected, it follows a linear decreasing dependency with the energy due to the constant peak time of the preamplifier. The measured trigger jitter does not pose a limitation on the performance of a PET [9], since a coincidence time window of 20 ns is established for normal operation of the PET ring [14].…”
Section: × 4 Pixel Matrix Measurements Resultsmentioning
confidence: 99%
“…As expected, it follows a linear decreasing dependency with the energy due to the constant peak time of the preamplifier. The measured trigger jitter does not pose a limitation on the performance of a PET [9], since a coincidence time window of 20 ns is established for normal operation of the PET ring [14].…”
Section: × 4 Pixel Matrix Measurements Resultsmentioning
confidence: 99%
“…However, increasing the voxel density in the FOV corresponds to reducing the statistics per voxel with the effect of increasing the image noise. A pixel size of 0.25 mm is chosen to fulfill the sampling requirement for an imaging detector presented in [16]: ΔrFWHM3 where Δ r is the chosen pixel size, FWHM is the full-width at half-maximum of the PSF of the scanner (~0.8 mm in the center of the FOV according to [2]).…”
Section: Discussionmentioning
confidence: 99%
“…Examples of these detectors are cadmium zinc telluride 44 and cadmium telluride. 45,46 Time of flight The use of ToF information in the reconstruction process converts the uniform probability of an event occurring somewhere within a line between detectors (line of response) to a Gaussian probability with full width at half maximum given by the time resolution between detectors. The width of the Gaussian probability is related to a reduced region in space where the annihilation takes place.…”
Section: Scintillation Crystalsmentioning
confidence: 99%