Background: The aim of this study was to explore the value of SMI compared with conventional ultrasonography for assessing hepatic arterial blood flow after pediatric liver transplantation. Methods: From March 2018 to November 2018, a total of 105 pediatric recipients with biliary atresia underwent liver transplantation in our hospital. Ultrasound examinations were performed at the bedside in the intensive care unit to check the patency of the blood flow in the hepatic allograft. CDI, PDI, cSMI, and mSMI were performed to assess the display, orientation, and distribution of the graft hepatic artery. Ultrasound examinations were performed by one radiologist, and the images were judged by two observers. Results: The median age, weight, and height of the recipients were 6.97 (5.92, 9.58) months, 6.50 (6.00, 7.80) kg, and 64.00 (62.00, 68.00) cm, respectively. The measure of kappa agreement was 0.902, 0.889, 0.882, and 0.882 for CDI, PDI, cSMI, and mSMI, respectively. HAT occurred in 7 pediatric recipients and was confirmed by CTA (computed tomography angiography) and surgery. The diagnostic performance of sensitivity, specificity, PPV (positive predictive value), NPV (negative predictive value), and accuracy were 100%, 92.86%, 50%, 100%, and 93.33% for CDI and 100%, 98.98%, 87.50%, 100%, and 99.05% for SMI. Conclusions: As an additional method to CDI, SMI can clearly show the distribution of hepatic arterial blood flow and provide more details, thereby markedly improving the diagnostic performance of postoperative HAT.