SU‐C‐211‐04: Micro‐CT Geometric Accuracy Phantom for Improved Fiducial Localization in Image‐Guided Needle Positioning Systems

Abstract: Purpose: To quantify and correct the geometric accuracy of an x‐ray microCT scanner to improve fiducial localization in image‐guided needle positioning systems. Methods: A geometric accuracy phantom was constructed by creating a one‐inch cube consisting of a 4×4×4 3D matrix of calibrated ¼” diameter spherical beads alternating between two materials. The phantom was imaged at both the isocenter and at the edge of the field of view of a micro‐CT scanner. All bead positions were measured in sets of five images an… Show more