SU‐D‐209‐02: Percent Depth Dose Curves for Fluoroscopic X‐Ray Beam Qualities Incorporating Copper Filtration

Abstract: Purpose: The purpose of this investigation was to quantify percent depth dose (PDD) curves for fluoroscopic x‐ray beam qualities incorporating added copper filtration. Methods: A PTW (Freiburg, Germany) MP3 water tank was used with a Standard Imaging (Middleton, WI) Exradin Model 11 Spokas Chamber to measure PDD curves for 60, 80, 100 and 120 kVp x‐ray beams with copper filtration ranging from 0.0–0.9 mm at 22cm and 42cm fields of view from 0 to 150 mm of water. A free‐in‐air monitor chamber was used to normal… Show more