Abstract:Purpose: To quantify the effect of pitch on noise in abdominal multi‐detector CT. Methods: The standard acrylic abdominal CTDI phantom was modified so that it could be supported on the gantry shroud and remain stationary during a helical multi‐detector CT image acquisition. The table top height was positioned 1 cm below the phantom so that the table could move during helical image acquisition without touching the phantom. The phantom was scanned with the x‐ray tube voltage, x‐ray tube anode current, reconstruc… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.