SU‐E‐I‐71: KVp Dependence of Transmitted Exposure for a Radiography Unit

Abstract: Purpose:To investigate the kVp dependence of the transmitted exposure for a radiography x‐ray unit.Methods:The study used a GE DiscoveryTM XR656 DR unit, a 30 (L) × 30 (W) × 25 cm thick Lucite phantom, two anthropomorphic phantoms (an Alderson RS‐310 chest phantom and a 3M skull phantom), an Unfors detector, and a Radcal 10×9‐6 ion chamber. We measured the entrance exposure and transmitted exposure of each phantom at 60, 70, 80, 90, 100, 110, 120 kVp for mAs range from 2.5 to 200 mAs, without any additional fi… Show more