Abstract:Purpose:To investigate the kVp dependence of the transmitted exposure for a radiography x‐ray unit.Methods:The study used a GE DiscoveryTM XR656 DR unit, a 30 (L) × 30 (W) × 25 cm thick Lucite phantom, two anthropomorphic phantoms (an Alderson RS‐310 chest phantom and a 3M skull phantom), an Unfors detector, and a Radcal 10×9‐6 ion chamber. We measured the entrance exposure and transmitted exposure of each phantom at 60, 70, 80, 90, 100, 110, 120 kVp for mAs range from 2.5 to 200 mAs, without any additional fi… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.