SU‐F‐P‐32: A Phantom Study of Accuracy of Four‐Dimensional Cone‐Beam CT (4D‐CBCT) Vs. Three‐Dimensional Cone Beam CT (3D‐CBCT) in Image Guided Radiotherapy

Abstract: Purpose: SymmetryTM 4D IGRT system of Elekta has been installed at our institution, which offers the 4D CBCT registration option. This study is to evaluate the accuracy of 4D CBCT system by using the CIRS 4D motion phantom and to perform a feasibility study on the implementation of 4D‐CBCT as image guidance for SBRT treatment. Methods: The 3D and 4D CT image data sets are acquired using the CIRS motion phantom on a Philips large bore CT simulator. The motion was set as 0.5 cm superior and inferior directions w… Show more