SU‐F‐P‐32: A Phantom Study of Accuracy of Four‐Dimensional Cone‐Beam CT (4D‐CBCT) Vs. Three‐Dimensional Cone Beam CT (3D‐CBCT) in Image Guided Radiotherapy
Abstract:Purpose:
SymmetryTM 4D IGRT system of Elekta has been installed at our institution, which offers the 4D CBCT registration option. This study is to evaluate the accuracy of 4D CBCT system by using the CIRS 4D motion phantom and to perform a feasibility study on the implementation of 4D‐CBCT as image guidance for SBRT treatment.
Methods:
The 3D and 4D CT image data sets are acquired using the CIRS motion phantom on a Philips large bore CT simulator. The motion was set as 0.5 cm superior and inferior directions w… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.