SU‐F‐T‐263: Dosimetric Characteristics of the Cine Acquisition Mode of An A‐Si EPID

Abstract: Purpose: To investigate the dosimetric characteristics of Varian a‐Si‐500 electronic portal imaging device (EPID) operated in cine mode particularly considering linearity with delivered dose, dose rate, field size, phantom thickness, MLC speed and common IMRT fields. Methods: The EPID that attached to a Varian Clinac 21iX linear accelerator, was irradiated with 6 and 18 MV using 600 MU/min. Image acquisition is controlled by the IAS3 software, Trigger delay was 6 ms, BeamOnDelay and FrameStartDelay were zero. … Show more