2016
DOI: 10.1118/1.4956757
|View full text |Cite
|
Sign up to set email alerts
|

SU-F-T-572: Small Fields Output Factors Measurement and Correction Factor Determination for Several Active and Passive Detectors

Abstract: Purpose: To measure the output factor (OF) and output correction factors (OCF) of small circular cones using various active and passive detectors. Methods: The OF of BrainLab circular cones of 4, 6, 7.5, 10, 12.5 and 15mm diameter were measured isocentrically at 15mm depth for 6 MV X‐rays at 1000 MU/min from Novalis‐Tx Linac using high resolution diodes (SFD, Edge, PTW60008), plastic scintillator (ExtradinW1), small ionization chamber (CC01), EBT3 film and Al2O3:C OSLDs (nanoDot). OCF was calculated as a ratio… Show more

Help me understand this report

This publication either has no citations yet, or we are still processing them

Set email alert for when this publication receives citations?