Abstract:Purpose:
To measure the output factor (OF) and output correction factors (OCF) of small circular cones using various active and passive detectors.
Methods:
The OF of BrainLab circular cones of 4, 6, 7.5, 10, 12.5 and 15mm diameter were measured isocentrically at 15mm depth for 6 MV X‐rays at 1000 MU/min from Novalis‐Tx Linac using high resolution diodes (SFD, Edge, PTW60008), plastic scintillator (ExtradinW1), small ionization chamber (CC01), EBT3 film and Al2O3:C OSLDs (nanoDot). OCF was calculated as a ratio… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.