Abstract:Purpose: Manufacturer recommended rounded leaf offsets for the Philips Pinnacle treatment planning system are inaccurate for leaves away from the central axis, and for varying energies. The need for an in‐house MLC rounded leaf offset correction is established, as well as a method for determining the offsets for individual MLC leaves. Materials and Methods: A standard picket‐fence test is developed with 2 cm separations across a 25×25 field. Dose distributions are calculated for the test, with 20th leaf offset… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.