Synchronous detection of vascular tension and nitric oxide release in pulmonary artery: A combined application of confocal wire myograph with confocal laser scanning microscopy
Abstract:Objectives To detect the vascular tension and nitric oxide (NO) release synchronously in mice pulmonary artery, we perform two experiments and present a novel application of confocal wire myograph coupled with the confocal laser scanning microscopy. Methods In the first experiment, viable endothelium-intact mouse pulmonary artery (outer diameter 100–300 μM) rings underwent a one-hour preincubation with a NO-specific fluorescent dye, 4-amino-5-methylamino-2′,7′-difluorofluorescein diacetate Calbiochem (2.5 μM),… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.