2020
DOI: 10.1016/j.hrcr.2020.05.003
|View full text |Cite
|
Sign up to set email alerts
|

Syncope after successful implantation of atrioventricular synchronous leadless pacemaker caused by polymorphic ventricular tachycardia

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
6
0

Year Published

2021
2021
2024
2024

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 5 publications
(6 citation statements)
references
References 16 publications
0
6
0
Order By: Relevance
“…Due to the dependency of the timing of A3 and A4 on heart rate, the device's delineation of A4 signals depends on an adequate parameter setting. Otherwise, AV synchrony may be perturbed and the risk of malignant arrhythmias might increase 13 . We consistently programmed the A3 window earlier (shortening of min.…”
Section: Critical Parameters For Av Synchronymentioning
confidence: 99%
“…Due to the dependency of the timing of A3 and A4 on heart rate, the device's delineation of A4 signals depends on an adequate parameter setting. Otherwise, AV synchrony may be perturbed and the risk of malignant arrhythmias might increase 13 . We consistently programmed the A3 window earlier (shortening of min.…”
Section: Critical Parameters For Av Synchronymentioning
confidence: 99%
“…The potential proarrhythmic effect of cardiac pacing is a rare but well-described phenomenon with LPs. 2 , 3 , 4 , 5 , 6 , 7 , 8 During the initial 140 implantations of the Micra transcatheter pacing system, Ritter and colleagues 2 reported 2 cases of ventricular tachycardia and 1 case of ventricular fibrillation. From a cohort of 92 Micra implants in Switzerland, Valiton and colleagues 4 reported a case of intractable VT after Micra implant.…”
Section: Discussionmentioning
confidence: 99%
“…In previously reported cases, “short-long-short” sequences that set the stage for “R-on-T”-induced ventricular arrhythmia and myocardial ischemia or scar have been proposed as potential causes of ventricular arrhythmia post LP implant. 3 , 6 , 8 …”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…In this case, the Micra-AV was initially implanted as a bridge therapy after biventricular pacemaker lead extraction, until a new transvenous biventricular system could be reimplanted after a complete resolution of the infectious process. Similarly, Halawa et al [20] reported a case of a polymorphic ventricular tachycardia caused by atrial undersensing during VDD pacing, that led to "short-long-short" sequences or "R-on-T" phenomenon. As the authors suggested, this event may show a relationship with specific Micra-AV algorithms sought to minimize ventricular pacing.…”
Section: Micra-av In a Real-life Settingmentioning
confidence: 94%