“…The lifetimes of all wells were averaged, obtaining mean values of and for Ribloflavin, PPIX and FAD, respectively. As the literature reports a broad range of lifetimes for autofluorescence, as seen in [33], individual lifetime characterization of the autofluorophores was also performed through a Time‐Domain Mesoscopic Fluorescence Imager (TD‐MFMT) [38] and Alligator‐based lifetime quantification [36]. One well per autofluorophore was measured from the same stock solutions and after lifetime calculation, linear regression of averaged lifetime values obtained with both systems yielded p‐value of 2.3e‐3, = 1, intercept = −0.23 and slope = 1.13, indicating high correlation between the estimated values for both systems, as shown in Appendix B (B) ‐ Figure A3.…”