2017
DOI: 10.1177/1457496916683093
|View full text |Cite
|
Sign up to set email alerts
|

Systematic Screening of Adverse Reactions to Metal Debris after Recap-M2A-Magnum Metal-on-Metal Total Hip Arthroplasty

Abstract: We found a high prevalence of adverse reaction to metal debris in the ReCap-M2a-Magnum total hip arthroplasty patients in this study; however, most of the patients did not require revision operations.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

0
12
0

Year Published

2018
2018
2023
2023

Publication Types

Select...
7

Relationship

0
7

Authors

Journals

citations
Cited by 10 publications
(12 citation statements)
references
References 21 publications
0
12
0
Order By: Relevance
“…Cr and Co ion measurements and Oxford Hip Score (OHS) questionnaires are checked on a 1-to 2-year interval depending on the symptom state and previous metal ion levels. The OHS questionnaire scale has been presented previously (14). If the patient had a symptomatic hip (poor or moderate postoperative OHS scores), or Cr and/or Co concentrations above 5 ppb, they were scheduled for MARS-MRI imaging of the hip.…”
Section: Methodsmentioning
confidence: 99%
“…Cr and Co ion measurements and Oxford Hip Score (OHS) questionnaires are checked on a 1-to 2-year interval depending on the symptom state and previous metal ion levels. The OHS questionnaire scale has been presented previously (14). If the patient had a symptomatic hip (poor or moderate postoperative OHS scores), or Cr and/or Co concentrations above 5 ppb, they were scheduled for MARS-MRI imaging of the hip.…”
Section: Methodsmentioning
confidence: 99%
“…In our previous report the 5-year survival of ReCap-M2a-Magnum THA was 93% for any reason for revision. 104 (8%) of 1,329 hips were revised after a median follow-up of 5 years, and ARMD was reported as the reason for revision in 33 hips (32% of revisions) ( 9 ). In the present study 197 hips (12%) had been revised and ARMD was the reason for revision in 50% of these reoperations at median of 10 years from the implantation.…”
Section: Discussionmentioning
confidence: 99%
“…Measurements were performed using the angle measurement tool of Carestream Vue PACS software ( www.carestream.com/en/ca/ ). The details of this screening program have been explained in our previous publications ( 9 , 12 , 13 ). Data was obtained from the Hospital District of Southwest Finland data pool and electronic medical records.…”
Section: Methodsmentioning
confidence: 99%
See 2 more Smart Citations