Technical Note: Field experiences using UV/VIS sensors for high-resolution monitoring of nitrate in groundwater

Abstract: Abstract. Two different in-situ spectrophotometers are compared that were used in the field to determine nitrate-nitrogen (NO3-N) concentrations at two distinct spring discharge sites. One sensor was a double wavelength spectrophotometer (DWS) and the other a multiple wavelength spectrophotometer (MWS). The objective of the study was to review the hardware options, determine ease of calibration, accuracy, influence of additional substances and to assess positive and negative aspects of the two sensors as well … Show more

Help me understand this report
View published versions

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

0
9
0

Year Published

2017
2017
2021
2021

Publication Types

Select...
5

Relationship

1
4

Authors

Journals

citations
Cited by 5 publications
(9 citation statements)
references
References 32 publications
0
9
0
Order By: Relevance
“…Attributes such as the type, source and purpose of these data are itemised in Table 1. In addition, data from a perennial spring on site were examined previously by Huebsch et al (2015) (Fig 1). In summary, a pump house and lagoon (23 m 2 ), is equipped with a Corbett trapezoidal flume used to measure discharge from the lagoon with a calibrated water level recorder (Diver; 0.000 m to 10.000 m).…”
Section: Secondary Datasetsmentioning
confidence: 99%
See 1 more Smart Citation
“…Attributes such as the type, source and purpose of these data are itemised in Table 1. In addition, data from a perennial spring on site were examined previously by Huebsch et al (2015) (Fig 1). In summary, a pump house and lagoon (23 m 2 ), is equipped with a Corbett trapezoidal flume used to measure discharge from the lagoon with a calibrated water level recorder (Diver; 0.000 m to 10.000 m).…”
Section: Secondary Datasetsmentioning
confidence: 99%
“…In summary, a pump house and lagoon (23 m 2 ), is equipped with a Corbett trapezoidal flume used to measure discharge from the lagoon with a calibrated water level recorder (Diver; 0.000 m to 10.000 m). A double wavelength spectrophotometer (5 mm path length) sensor (NITRATAX plus sc, Hach Lange GmbH, Germany; Huebsch et al, 2015) measured NO 3 -N concentrations (0.000 to 25.000 mg L -1 ). Both instruments measured at a 15 minute resolution and were used to calculate NO 3 -N flux from 27 November 2011 to 25 March 2013.…”
Section: Secondary Datasetsmentioning
confidence: 99%
“…multiwavelength one (MWS) with PLSR resolution was carried out by Huebsch et al (2015) for groundwater monitoring. The findings were that the MWS offers more possibilities for calibration and error detection, but requires more expertise compared with the DWS.…”
Section: Accepted Manuscriptmentioning
confidence: 99%
“…4 In addition, broad-band ultraviolet/visible spectroscopy-based on-line probes for water monitoring have emerged. 5 These systems have been employed for complex water streams, e.g. with high turbidity, relying on sophisticated calibrations dedicated to each application.…”
Section: Introductionmentioning
confidence: 99%
“…with high turbidity, relying on sophisticated calibrations dedicated to each application. [5][6][7] Apart from these established techniques, infrared (IR) spectroscopy provides quantitative chemical information that can be obtained in reagent free measurements in real time. Ions like nitrate show strong and specific bands in the mid-IR spectrum.…”
Section: Introductionmentioning
confidence: 99%