Technical Note: Imaging dose resulting from optimized procedures with limited‐angle intrafractional verification system during stereotactic body radiation therapy lung treatment
Abstract:Purpose
The limited‐angle intrafractional verification (LIVE) system was developed to track tumor movement during stereotactic body radiation therapy (SBRT). However, the four‐dimensional (4D) MV/kV imaging procedure results in additional radiation dose to patients. This study is to quantify imaging radiation dose from optimized MV/kV image acquisition in the LIVE system and to determine if it exceeds the American Association of Physicists in Medicine Task Group Report 180 image dose threshold.
Methods
TrueBea… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.