Abstract:Although no fluorescence was detected, the mechanical operation of the FLD was satisfactory throughout 8 hours of flight, with attendant vibrations, and during a wide range of sun angles, from 13° to 33°. In accord with design, the FLD showed no appreciable response to moderate contrasts in reflectivity among targets, although there was a marked response to strong contrasts in reflectivity. Airborne tests also established that a minimum instrumental warm-up time of seven minutes is required, and that response … Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.