Background Few studies have examined factors that affect acetabular cup positioning. Since cup positioning has been linked to dislocation and increased bearing surface wear, these factors affecting cup position are important considerations. Question/purposes We determined the percent of optimally positioned acetabular cups and whether patient and surgical factors affected acetabular component position. Methods We obtained postoperative AP pelvis and crosstable lateral radiographs on 2061 consecutive patients who received a THA or hip resurfacing from 2004 to 2008. One thousand nine hundred and fifty-two hips had AP pelvic radiographs with correct position of the hip center, and 1823 had both version and abduction angles measured. The AP radiograph was measured using Hip Analysis Suite TM to calculate the cup inclination and version angles, using the lateral film to determine version direction. Acceptable ranges were defined for abduction (30°-45°) and version (5°-25°).Results From the 1823 hips, 1144 (63%) acetabular cups were within the abduction range, 1441 (79%) were within the version range, and 917 (50%) were within the range for both. Surgical approach, surgeon volume, and obesity (body mass index [ 30) independently predicted malpositioned cups.