The Effectiveness of Ultraviolet Smart D60 in Reducing Contamination of Flexible Fiberoptic Laryngoscopes

Abstract: ObjectiveTo compare the effectiveness of disinfection protocols utilizing a ultraviolet (UV) Smart D60 light system with Impelux™ technology with a standard Cidex ortho‐phthalaldehyde (OPA) disinfection protocol for cleaning flexible fiberoptic laryngoscopes (FFLs).MethodsTwo hundred FFLs were tested for bacterial contamination after routine use, and another 200 FFLs were tested after disinfection with one of four methods: enzymatic detergent plus Cidex OPA (standard), enzymatic detergent plus UV Smart D60, mi… Show more