Abstract:Purpose: The purpose of this study was to investigate the impact of a new arterial intravascular pump on the hemodynamic surroundings within the aorta. Methods: A new arterial intravascular pump was placed in the descending aorta, and the effects of three positions within the aorta, as well as the number (n = 1 to 3) of pumps, on arterial flow features, organ perfusion, and blood trauma were investigated using a computational fluid dynamics (CFD) method. Results: It was found that as the pump position was move… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.