Background
This study's intent is to evaluate the usefulness of pattern matching filter (PMF) function combined with robotic magnetic navigation (RMN) in guiding the ablation of premature ventricular contractions (PVCs).
Hypothesis
Assume that PMF can improve the outcomes of PVCs ablation using RMN.
Methods
A retrospective analysis was completed consisting of 118 consecutive patients with PVCs who underwent radiofrequency ablation guided by RMN. According to the application of PMF, patients were divided into two groups: 20 patients underwent ablation without PMF (group A), and another 98 patients received ablation incorporating PMF (group B).
Results
Compared with group A, the procedure time (135.0 ± 28.3 min vs. 106.3 ± 37.9 min, p = 0.02) in group B was significantly decreased, while the X‐ray exposure time (6.0 ± 2.6 min vs. 6.5 ± 3.6 min, p = 0.705) and dose (3.2 ± 2.4 gycm2 vs. 3.9 ± 2.7 gycm2,p = 0.208) had no significant difference. Group B had a more than twofold number of points acquired (66.9 ± 23.0 vs. 143.9 ± 68.3, p < 0.001) and required a shorter radiofrequency ablation time (13.2 ± 3.5 min vs. 8.1 ± 2.9 min, p < 0.001). There were no serious complications in either group. The acute success rate was similar [90.0% (18/20) vs. 87.8% (86/98), p = 1.000] in two groups, and the success rate was also similar in the long‐term follow‐up [83.3% (15/18) vs. 87.2% (75/86), p = 0.776].
Conclusions
The ablation of PVCs guided by RMN is safe and effective. Combined with the functional capability of PMF, both procedure time and radiofrequency ablation time were significantly decreased.